Accurate, Reliable Sodium Monitoring
Combination sodium and pH electrode system confirms reagent addition for reliable operation, minimizes reagent consumption, and reduces maintenance.
Configurable, Automated Calibration
Unattended, automated calibration ensures the accurate measurement of sodium in your water system and detects breakthroughs right when they occur.
Fully Enclosed for Dirty Environments
Especially suitable for plant floor locations in coal-fired power plants and where a locking door is needed.
Material No.: 58042002
Sodium Measurement Range | 0.001 - 100,000 ppb or equivalent ppm, autoranging |
Sodium Measurement Accuracy | ±10% of reading or ± 0.05 ppb whichever is greater |
Response Time | 90% at 5 minutes |
Sample pH | 2.5-12 |
Update Rate | Once per second |
Reagent Consumption | Diisopropylamine (DIPA), approximately 1 L per 3 months more at higher temperatures and for cation exchange samples |
Sample Flow rate | >40 mL/min (>20 mL/min for cation exchange samples), excess to drain |
Sample Temperature | 5-50°C (41-122°F) |
Sample Pressure | 0.3-7 bar (5-100 psig) |
Grab Sample Measurement | Included |
Power Supply | 100-240 VAC, 50-60 Hz, 25 W, on power loss all settings are retained without batteries |
Dimensions (HxWxD) | 35.4"x17.7"x7.5" (900x450x190 mm) |
Weight | 27 kg (60 lbs) |
Ambient Operating Temperature | 10-45°C (50-113°F) |
Humidity | 10-90% non-condensing |
Process Connection | Sample Inlet: 1/4" or 6 mm OD tube SS compressin fitting Drain Hose: 19x25.4 mm (3/4 x 1") 2 m (6 ft) length included |
Operator Interface | 4-line backlit LCD, 5 tactile keys simultaneous display of sodium, conditioned pH, auto-cal status (temperature optional) |
Analog Output | For sodium, conditioned pH, temperature four powered 0/4-20 mA, 22mA alarm, 500 ohm max load, not for use with externally powered circuit |
Analog Output Scaling | Linear, bi-linear, logarithmic (1,2,3,4 decades), auto ranging |
Measurement Parameter | Sodium |
Electrode Conditioning | Part of auto-cal sequence |
Relay Contacts | 3A resistive freely assignable to setpoints for sodium, pH, temperature other relays used for auto-cal Two unpowered, SPDT, 250VAC/30VDC |
Sodium Calibration | Automatic, unattended 3-point known addition |
Calibration | Automatic |
Analog Output Accuracy | ±0.05 mA |
Model Family | 2300Na |
EnclosureType | Full |
Built-in sequencer | No |
Sub-PPB Sodium Monitoring at a Glance
The 2300Na Sodium Analyzer delivers more than monitoring at sub-ppb sodium concentration levels. The simultaneous display of temperature, adjusted pH and calibration progress provide a complete picture to increase uptime and reduce operating costs.
Greater Accuracy (Unattended)
The 2300Na Sodium Analyzer eliminates the need for operator intervention with pre-configured, periodic self-calibration. Providing greater measurement accuracy, an internal pH measurement ensures interference-free sodium determination.
Full Protection in Dirty Environments
The fully enclosed 2300Na Sodium Analyzer protects all components, preventing damage from outside interference. This ensures safe, reliable and continuous operation in production environments.
Low Maintenance Made Easy
With 6-month preventive maintenance intervals, the 2300Na Sodium Analyzer requires less downtime and reduces the overall critical operator time commitment. Every calibration automatically results in sensor conditioning, further reducing maintenance requirements.